Description
This package can be used to communicate a SolsTiS M Squared CW
Ti-saphire laser using TCP/IP. The functions can be used to send a
general command, or use the built in functionality for getting and
setting the wavelngth.
The main nontrivial trick was to count '}' characters instead of
waiting for a terminator. The Solstis does not send a terminator
character over TCP/IP, so if Matlab expects one, then there is a long
lag until the timeout is reached.
The SolsTiS laser should be set up in a standard way using an
Angstrom wavemeter. Setup involves setting the TPC/IP addresses
correctly in the web interface. In the solstis interface, set the
remote interface to the control computer's ip address. Then disable
and enable the link. You can find a windows machine's ip address by
typing ipconfig into the command prompt.
Functions.
solstis_findInstrument - initialize solstis instrument object
solstis_getWavelength - read wavelength
solstis_gotoWavelength - go to wavelength
solstis_clearBuffer - Clear all incoming messages
solstis_getResponse - read solstis instrument
sostis_query - query solstis instrument
solstis_processResponse - processes solstis text response
Zitieren als
Todd (2026). solstislab (https://de.mathworks.com/matlabcentral/fileexchange/48669-solstislab), MATLAB Central File Exchange. Abgerufen .
Quellenangaben
Inspiriert: Solstis class, Controllers for Thorlabs PM100 (2006/older era)
Kategorien
Mehr zu Instrument Connection and Communication finden Sie in Help Center und MATLAB Answers
Allgemeine Informationen
- Version 1.0.0.0 (7,85 KB)
Kompatibilität der MATLAB-Version
- Kompatibel mit allen Versionen
Plattform-Kompatibilität
- Windows
- macOS
- Linux
| Version | Veröffentlicht | Versionshinweise | Action |
|---|---|---|---|
| 1.0.0.0 |